Time Base Accuracy; Description; Clock Verification Procedure - LeCroy WaveSurfer 400 Series Service Manual

Color digital oscilloscopes
Table of Contents

Advertisement

5.11 Time Base Accuracy

5.11.1 Description

An external sine wave generator of 10.0 MHz with frequency accuracy better than 1
PPM is used.
Specifications
Clock : accuracy : ≤ ±0.001 % or ≤ ±10 PPM
5.11.2
5-34 Performance Verification

Clock Verification Procedure

Recall Timebase Accuracy.lss or configure the DSO
Panel Setups
:
Channels trace ON
Input gain
:
Input offset
:
Trigger setup
:
Trigger on
:
Slope 1
:
Level 1
:
Trigger mode
:
Delay
:
Time base
:
Measure
:
P1
:
Connect the RF output of the HP8648B generator through a 50 Ohm
coaxial cable into Channel 1.
Set the generator frequency to 10.0 MHz.
Adjust the generator output amplitude to get 6 divisions peak to peak.
Read of the frequency parameter (to 2 decimal places) and record the value in
Table 14.
Verify that the error is less than 100 Hz.
Recall FROM DEFAULT SETUP
Channel 1
.1 V/div.
0 mV
Edge
C1
Pos
0 mV
Auto
50 %
100 msec/div.
Parameters
Frequency of C1
Rev. D Sept 2007