LeCroy WaveSurfer 400 Series Service Manual page 64

Color digital oscilloscopes
Table of Contents

Advertisement

Ref output.
Sensor Input
5-24 Performance Verification
Recall Bandwidth - CH1 10mv.lss or configure the DSO:
Panel Setups
:
Channels Trace ON
Input Coupling
:
Input gain
:
Input offset
:
Trigger setup
:
Trigger on
:
Mode
:
Time base
:
Change parameters
P1
:
P2
:
Connect the HP8482A power sensor to the power meter.
Zero and calibrate the HP8482A power sensor using the power meter Power
Connect a BNC adapter to the HP8482A power sensor.
Connect a 50Ω SMA cable to the RF output of the HP8648B
generator and then through the necessary adapters to the power sensor. It is
very important that the same cable/generator be used throughout this BW
procedure and that the SMA connectors are torqued at all their mating locations.
Power Meter
Power Ref
Output
Power Sensor
Figure 5-3 : Power Meter Equipment Setup
Set the generator frequency to 10 MHz
Set the generator amplitude to measure 8 µW on the power meter.
Read the displayed generator output amplitude, and record it in the third
column of Table 9.
Recall FROM DEFAULT SETUP
Channel 1
DC 50Ω on all 4 Channels
10 mV/div on all 4 Channels
0 mV on all 4 Channels
Edge
Line
Auto
50 ηsec/div.
PK-PK of C1
Freq of C1
Sine Wave
Generator
Rev. D Sept 2007