Performance Verification; Introduction; List Of Tested Characteristics; Calibration Cycle - LeCroy WaveSurfer 400 Series Service Manual

Color digital oscilloscopes
Table of Contents

Advertisement

5.

Performance Verification

5.1

Introduction

This chapter contains procedures suitable for determining if the WaveSurfer 400
Series of Digital Storage Oscilloscope performs correctly and as warranted. They
check all the characteristics listed in subsection 5.1.1.
In the absence of the computer automated calibration system based on LeCroy
Calibration Software (Calsoft), this manual performance verification procedure can
be followed to establish a traceable calibration. It is the calibrating entities'
responsibility to ensure that all laboratory standards used to perform this procedure
are operating within their specifications and traceable to required standards if a
traceable calibration certificate is to be issued for the WaveSurfer 400 series Digital
Storage Oscilloscope.

5.1.1 List of Tested Characteristics

This subsection lists the characteristics that are tested in terms of quantifiable
performance limits.
• Input Impedance
• Leakage Current
• Peak to Peak noise level
• Positive and Negative DC accuracy
• Positive and Negative Offset
• Bandwidth
• Trigger Accuracy
• Time Base Accuracy

5.1.2 Calibration Cycle

The WaveSurfer 400 series Digital Storage Oscilloscope requires periodic
verification of performance. Under normal use (2,000 hours of use per year) and
environmental conditions, this instruments calibration cycle is 12 months.
Rev. D Sept 2007
Performance Verification 5-1