10. H O T A I R G E N E R AT O R
Problem
Possible cause/remedy
No aerosol is formed
Check whether the set and actual temperatures are
the same (at least 180°C (356°F)).
Check whether the Filling Chamber and/or Valve
Balloon or Tube are correctly attached.
Check whether dronabinol has dropped onto the Drip
Pad or whether there is fresh, unused cannabis floss
in the Filling Chamber.
The Valve Balloon does
Check whether the pump is turned on.
not fill up.
Check whether the Filling Chamber and/or Valve
Balloon are correctly attached.
The device switches off
The automatic switch off has intervened. Switch the
(automatic switch-off).
heater or pump on again.
Plant particles in the Val-
Check whether the Screen in the Filling Chamber Cap
ve Balloon or Tube (when
is clogged or is not inserted into the notch properly.
using cannabis flos).
Check whether the Filling Chamber was cleaned
according to the Instructions for Use.
60
10. H O T A I R G E N E R AT O R
If the preceding tips do not
yield the desired results, or
for any other defects or prob-
lems not mentioned here, please pull
the power plug and contact our Ser-
vice Center immediately.
Do not open the Hot Air Gen-
erator! Without special tools
and specialized knowledge,
any attempt to open the device will
cause damage to it. Such an attempt
will invalidate the warranty claim.
Do not try to remedy the defect your-
self. Contact directly our Service
Center.
10.7. Technical Specifications
VOLCANO MEDIC 2

Vaporization System

Voltage:
220-240 V / 50-60 Hz (alternatively
available with 110-120 V / 50-60 Hz)
The voltage specification is located
on the bottom side of the
VOLCANO MEDIC 2 Hot Air Generator.
Power consumption: 270 VA
Air pump output: 10 VA
Air temperature: adjustable between
approx. 40 - 210°C (104 - 410°F)
Air flow: approx. 12 l/min
Valve Balloon Size:
approx. 12.5 liters
Size of the Hot Air Generator:
20.0 x 18.0 cm (7.9 x 7.1 inch)
Weight of the Hot Air Generator with
Power Cord, Filling Chamber, and
Valve Balloon with Mouthpiece:
approx. 1.8 kg (3.9 lbs)
Safety category II
Automatic switch off after 30 minutes
Please unplug to disconnect from the
mains current.
Internal fuse F1: F-T 4 A (applies to
110-120 V and 220-240 V devices)
This fuse may be replaced only by our
Service Center.
Subject to technical changes.
Patents and designs:
www.vapormed.com/patents
Manufacturer:
Storz & Bickel GmbH & Co. KG
In Grubenäcker 5-9,
78532 Tuttlingen, Germany
61