6 . T H E U S E O F D R O N A B I N O L
Caution! Hot surface!
Do not touch any components
of the Filling Chamber (except
at the textured grip) until they
have cooled down after the
Valve Balloon has been filled.
Separate the Valve Balloon from the
Filling Chamber. The Valve closes au-
tomatically after separation and the
aerosol in the Valve Balloon cannot
escape.
Detach Valve Balloon from Filling Chamber
Do not leave the Filling Cham-
ber on the hot air outlet of the
VOLCANO MEDIC 2 when the heater is
switched on – except for separating
the alcohol or filling the Valve Balloon.
Caution! Hot surface!

6.2.3. Inhalation

Attach the Mouthpiece to the Valve and
allow it to lock in place.
Attach the Mouthpiece onto the Valve
Balloon
32
6 . T H E U S E O F D R O N A B I N O L
Correct position of the Mouthpiece on the
Valve
To inhale, put the Mouthpiece to your
lips and lightly press against the
Mouthpiece with your lips. By doing so,
the Valve opens and the aerosol can
be inhaled from the Valve Balloon.
As soon as you remove the pressure
from the Mouthpiece, the Valve auto-
matically closes.
Press Mouthpiece against lips to inhale
To avoid residual moisture
deposition in the Valve Bal-
loon, and to minimize the risk
of possible contamination, do
not blow into the Valve Bal-
loon.
As the content of the Valve Bal-
loon deflates, it may be com-
pletely emptied by stretching
the Valve Balloon taut and subse-
quently inhaling the content.
Stretch Valve
Balloon tight
The Valve Balloon is not suit-
able for storing the aerosol
for long periods, as it will con-
dense on the interior surface of the
Balloon in time (a few hours).
Inhale the contents of the Valve Bal-
loon within 10 minutes.