Storage; In Case Of Problems; Disposal; Troubleshooting - Vapormed Volcano MEDIC 2 Instructions For Use Manual

Hide thumbs Also See for Volcano MEDIC 2:
Table of Contents

Advertisement

10. H O T A I R G E N E R AT O R
Reassemble in
reverse order

10.3. Storage

Store the Hot Air Generator in a dry
place protected against the effects of
weather and out of the reach of chil-
dren or unqualified persons.

10.4. In Case of Problems

Immediately unplug the Power Cord
from the mains supply.
If there are any faults in the electron-
ics of the VOLCANO MEDIC 2 Hot Air
Generator, you can perform a reset by
pulling out the power plug and plug-
ging it again after waiting for minimum
three seconds, or by keeping the touch
button labeled 'HEAT'' pressed for 10
seconds.
The Hot Air Generator has one internal
fuse (F1). As this fuse is used only in
case of a defect in the Hot Air Genera-
tor, it can be replaced only by our Ser-
vice Center.
Put the Hot Air Generator in the origi-
nal packaging or pack it properly in
other packing material and take it to
our Service Center.

10.5. Disposal

If the Hot Air Generator is irreparably
damaged, please do not simply throw
it away.
As it is made of high quality, complete-
ly recyclable components, it should be
brought to a material recycling centre
in your town or sent to our Service
Center for disposal.
58
10. H O T A I R G E N E R AT O R

10.6. Troubleshooting

If the device does not function properly when the power plug is connected to the
power socket, please check the Hot Air Generator using the tips in the following
table:
Problem
Possible cause/remedy
The VOLCANO MEDIC 2
Please make sure that the mains supply current is on
does not show mains op-
and that the fuse (in the mains fuse box) is switched
eration on the display; the
on and is not defective.
heater and pump cannot
If the Hot Air Generator still does not function, the
be activated.
internal fuse may have been triggered.
In this case, send the VOLCANO MEDIC 2 Hot Air
Generator to our Service Center.
Message on the display:
Device faulty. Perform a factory reset if number + "r"
is displayed. Display of number + "n" is an irreversible
Line 1: "ERR"
error; the VOLCANO MEDIC 2 Hot Air Generator must
Line 2: number, followed
be sent to our Service Center.
by the letter "r" or "n"
Possible fault messages::
01r
Internal communication fault. Restart recommended.
02r
The device is either outside its specified temperature,
or internal overheating has occurred. Switch the
device off, pull the power plug out of the socket, and
leave the device for min. 30 mins at room tempera-
ture without operating it. If the fault continues after
30 mins., contact our Service Center.
03r
The device is being operated at the incorrect voltage.
Please check which version (115V/230V) of the
device you have, and ensure that you are using it at
the the correct voltage.
04n
Device is faulty, please send it to our Service Center.
59