Repair and maintenance

Maintenance and care

Cleaning intervals
For the sake of personal, environmental, and material protection, it is your duty to clean and if necessary,
disinfect the centrifuge on a regular basis.
Maintenance
Clean rotor chamber
Clean rotor
Accessories
Cabinet
Filter unit
Ventilation holes
CAUTION Refrain from using any other cleaning or decontamination procedure than those
recommended here, if you are not entirely sure that the intended procedure is safe for the
equipment.
Use only approved cleansers.
If in doubt, contact VWR.
Cleaning
When cleaning centrifuge and accessories mind the following:
Use warm water with a neutral solvent.
Never use caustic cleaning agents such as soap suds, phosphoric acid, bleaching solutions or
scrubbing powder.
Rinse the cavities out thoroughly.
Use a soft brush without metal bristles to remove stubborn residue.
Afterwards rinse with distilled water.
Place the rotors on a plastic grate with their cavities pointing down.
If drying boxes are used, the temperature must never exceed 50 °C, since higher temperatures
could damage the material and shorten the lifetime of the parts.
Use only disinfectants with a pH of 6-8.
CAUTION Before using any cleaning or decontamination methods except those recommended
by the manufacturer, users should check with the manufacturer that the proposed method will
not damage the equipment.
35
Recommended interval
Daily or when polluted
Daily or when polluted
Daily or when polluted
Once per month
Every three months
Every six months
VWR MICRO STAR 17 / 17R ver c, rel 09.06.2020