VWR MICRO STAR 17 Instruction Manual page 19

Hide thumbs Also See for MICRO STAR 17:
Table of Contents

Advertisement

The relative centrifugal force (RCF) is given as a multiple of the force of gravity g. It is a unitless
numerical value which is used to compare the separation or sedimentation capacity of various
centrifuges, since it is independent of the type of device. Only the centrifuging radius and the speed
RCF = 11.18 x (
come into play in it:
����
1000
r = centrifuging radius in cm
n = rotational speed in rpm
The maximum RCF value is related to the maximum radius of the tube opening.
Remember that this value is reduced depending on the tubes and adapters used.
This can be accounted for in the calculation above if required.
Run time preselection
Note: You can select a run time between 1 and 99 min or continuous operation.
1. Enter the desired value by pressing the key
the desired value shows.
You can adjust the run time in steps of 1 min.
2. Press the
key to confirm the preselected value.
If you do not press any key, the display flashes for a few seconds the new preselected value is now
stored, and the display shows the actual value.
Continuous operation
1. Press the
2. During continuous operation, the centrifuge will continue running until you stop it manually with
the
key.
CAUTION Please note that the lifetime of particularly plastic rotor tubes is limited. Continuous
operation (extended use) may cause damage to them.
19
Explanation of RCF value
)
x r
2
key until hd is shown.
below the display in the middle repeatedly, until
VWR MICRO STAR 17 / 17R ver c, rel 09.06.2020

Advertisement

Table of Contents
loading

This manual is also suitable for:

Micro star 17r521-1646521-1647