Clean Air Filter

Clean the Condenser

VWR International

Maintenance

Section 5
Warning Avoid the excessive use of water around the control area due to
the risk of electrical shock. Damage to the controls may also result.
Wipe down the freezer exterior using soap and water and a general use
laboratory disinfectant. Rinse thoroughly with clean water and dry with a
soft cloth.
Clean the air filter a minimum of four times per year.*
1. Open the front lower door by grasping the bottom left corner.
2. Locate the grille on the door. See Figure 1-5. Grasp the middle of the
grille material and gently pull out to remove.
3. Wash the filter material using water and a mild detergent.
4. Dry by pressing between two towels.
5. Install the filter back into the grille and attach the grille.
* The clean filter alarm occurs every three months as a reminder to clean the
air filter. Depending upon environmental conditions, the filter may need to be
cleaned or replaced more frequently. If the filter becomes torn or excessively
dirty, a replacement can be purchased from VWR. Order part number
760203.
Clean the condenser a minimum of once a year.*
1. Open the front lower door by grasping the bottom left corner. See
Figure 1-5.
2. Using a vacuum cleaner, exercising care to not damage the condenser
fins, clean the condenser.
* Depending upon environmental conditions, the condenser may need to be
cleaned more frequently.