Follow the instructions given in the rotor manual.
Never overload the rotor. Follow the instructions given in the rotor manual.
Never open the lid until the rotor has come to a complete stop and this has been
confirmed in the display.
The lid emergency release may be used in emergencies only to recover the samples
from the centrifuge, e.g. during a power failure.
Never use the centrifuge if parts of its cover panels are damaged or missing.
Do not touch the electronic components of the centrifuge or alter any electronic or
mechanical components.
Please observe the safety instructions.
Please pay particular attention to the following aspects:
Location: Well-ventilated environment, set-up on a level and rigid surface with
adequate load-bearing capacity.
Rotor installation: Make sure the rotor is locked properly into place before operating
the centrifuge.
Especially when working with corrosive samples (salt solutions, acids, bases), the
accessory parts and vessel have to be cleaned carefully.
Always balance the samples.
Centrifuging hazardous substances:
Do not centrifuge explosive or flammable materials or substances which could react
violently with one another.
The centrifuge is neither inert nor protected against explosion. Never use the
centrifuge in an explosion-prone environment.
Do not centrifuge inflammable substances.
Remaining risk: Improper use can cause damages, contamination and injuries
with fatal consequences.
Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms
without suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the
"Bio-safety Manual" of the World Health Organization (WHO)), aerosol-tight biological
seals have to be used.
For materials in a higher risk group, extra safety measures have to be taken.
If toxins or pathogenic substances have gotten into the centrifuge or its parts,
appropriate disinfection measures have to be taken.
Remaining risk: Improper use can cause damages, contamination and injuries
with fatal consequences.
Highly corrosive substances which can cause material damage and impair the
mechanical stability of the rotor, should only be centrifuged in corresponding protective
tubes.
8
VWR MICRO STAR 17 / 17R ver c, rel 09.06.2020