(a)
(b)
3.
Let the rotor slide slowly down the centrifuge spindle.
CAUTION Do not force the rotor onto the centrifuge spindle.
If the rotor is very light, then it may be necessary to press it onto the centrifuge spindle with a
bit of pressure.
4.
Thread the fastening tool into the centrifuge spindle clockwise. Hold the rotor with the other hand
into position.
5.
Close the rotor.
6.
Check if the rotor is properly installed by lifting it slightly on the handle.
WARNING Check for any damage to the rotor: Damaged rotors must not be used. Keep the
centrifuge spindle area of the rotor clear of objects.
CAUTION Check that the rotor is properly locked on the centrifuge spindle before each use by
pulling it at its handle. Fasten the rotor if necessary.
Be sure to check all seals before starting any aerosol-tight applications.
See the information in the rotor instruction manual.