Failure message
Display oP appears
although lid is closed.
Lid
bAL
(only VWR Micro Star)
E-01 - E-13
E-14
E-15 - E-16
30
Problem with centrifuge
Centrifuge does not start.
Rotor stops with deceleration to
standstill.
Rotor stops with deceleration to
standstill.
Rotor stops with deceleration to
standstill. The centrifuge cannot
be operated.
Rotor stops with deceleration to
standstill. The centrifuge cannot
be operated.
Rotor stops with deceleration to
standstill. The centrifuge cannot
be operated.
Possible causes and cures
Centrifuge door not properly closed.
Open the centrifuge door and repeat
locking procedure.
If the message appears again, inform
Customer Service.
Centrifuge door was opened manually
during the run. Close centrifuge door
immediately.
Rotor stops with deceleration to
standstill.
For further centrifugation, you have to
switch the instrument off and switch it
on again.
Imbalance switch releases.
1. Open the centrifuge by
pressing the
2. Check whether the rotor is
properly loaded.
3. Check whether a broken tube
or damaged rotor released the
imbalance switch.
If an error message appears
again, inform Customer
Service.
Internal program error
Switch the instrument off and on again.
If an error message appears again,
inform Customer Service.
Overtemperature in the centrifuge
chamber.
Switch the centrifuge off and turn it on
again after approx. one minute.
If an error message appears again,
inform Customer Service.
Temperature measurement error
Switch the instrument off and on again.
If an error message appears again,
inform Customer Service.
VWR MICRO STAR 17 / 17R ver c, rel 09.06.2020