Troubleshooting; Problems During A Firmware Update - R&S NRPM User Manual

Ota power measurement solution
Table of Contents

Advertisement

®
R&S
NRPM
Do not apply a signal to the sensor module while the selftest is running.
Executing the self-test with an active signal assigned, can erroneously result
in error messages at the test steps Offset Voltages and/or Noise Voltages.
Using remote control
► For a quick check, send TEST:SENSor?.
For each test step, PASS or FAIL is listed.
Using the web browser-based user interface (R&S NRPM3N LAN sensor
module)
1. In the navigation pane of the main dialog, select "Sensor".
2. Under "Diagnostics", select "Selftest".
See also
"Selftest"
12.3

Problems during a firmware update

The firmware update is described in
Solutions for potential problems that can occur when using the firmware update
for R&S NRP family, see
Firmware update was interrupted
If, for example, a power cut happened during the firmware update, problems can
occur.
1. Perform the firmware update again. Sometimes, a further update fixes the
problems.
2. If you nevertheless can not access the sensor module, contact the local ser-
vice representative.
User Manual 1425.8663.02 ─ 07
on page 79.
Chapter 8, "Firmware
"Firmware update for R&S NRP family"
Problems during a firmware update
update",

Troubleshooting

on page 81.
on page 82.
185