R&S
FSMR3000
13 Troubleshooting
If the results do not meet your expectations, the following sections may contain helpful
hints and information.
●
Error
●
Error Messages in Remote Control
●
Troubleshooting Remote
●
Miscellaneous Troubleshooting
●
System
●
Collecting Information for
●
Contacting customer
13.1 Error Information
If errors or irregularities are detected, a keyword and an error message, if available,
are displayed in the status bar.
Depending on the type of message, the status message is indicated in varying colors.
Table 13-1: Status bar information - color coding
Color
Red
Orange
Gray
No color
Green
If any error information is available for a channel, an exclamation mark is displayed
next to the channel name ( ). This is particularly useful when the MultiView tab is dis-
played, as the status bar in the MultiView tab always displays the information for the
currently selected measurement only.
Furthermore, a status bit is set in the STATus:QUEStionable:EXTended:INFO reg-
ister for the application concerned. Messages of a specific type can be queried using
the SYSTem:ERRor:EXTended? command. For more information, see the
R&S FSMR3 User Manual.
User Manual 1179.0116.02 ─ 02
Information................................................................................................... 692
Operation...................................................................... 694
Recovery.................................................................................................. 696
Support.........................................................................696
support................................................................................ 698
Type
Description
Error
An error occurred at the start or during a measurement, e.g. due to missing
data or wrong settings, so that the measurement cannot be started or com-
pleted correctly.
Warning
An irregular situation occurred during measurement, e.g. the settings no lon-
ger match the displayed results, or the connection to an external device was
interrupted temporarily.
Information
Information on the status of individual processing steps.
No errors
No message displayed - normal operation.
Measurement
Some applications visualize that the measurement was successful by show-
successful
ing a message.
Mode.............................................................693
Hints................................................................... 695
Troubleshooting
Error Information
692