Optimizing And Troubleshooting The Measurement - R&S FSW Series User Manual

Signal and spectrum analyzer
Hide thumbs Also See for FSW Series:
Table of Contents

Advertisement

®
R&S
FSW
Figure 7-38: Measurement of the R&S FSW's intrinsic noise power in a 1.23 MHz channel band-

7.2.8 Optimizing and Troubleshooting the Measurement

If the results do not meet your expectations, or if you want to minimize the measure-
ment duration, try the following methods to optimize the measurement:
Only activate as many adjacent channels as necessary to minimize the required
span and thus the required measurement time for the measurement.
Increase the RBW to minimize the measurement time; however, consider the
requirements of the standard if you need to measure according to standard! The
automatic settings are always according to standard.
Take advantage of the speed optimization mode in the "Sweep" settings if you do
not require the larger dynamic range (see
Reduce the "Sweep Time" and thus the amount of data to be captured and calcu-
lated; however, consider the requirements regarding the standard deviation.
To improve the stability of the measured results, increase the "Sweep Time",
which also leads to more averaging steps.
Instead of trace averaging, use an RMS detector with a higher "Sweep Time" to
obtain better average power results in less time.
To determine a channel power level quickly, use the
surement
ment is a zero span measurement where the sweep time determines the measure-
ment time. Due to the FFT measurement, duplicate averaging is performed, provid-
ing very stable results very quickly.
Note, however, that for TDP measurements, channel filters are not available and a
fixed RBW is used. Thus, the measurement may not be according to standard for
some test cases.
User Manual 1173.9411.02 ─ 43
Channel Power and Adjacent-Channel Power (ACLR) Measurement
width.
(TDP) rather than a Channel Power measurement. The TDP measure-
Measurements and Results
"Optimization"
on page 470).