Dräger Babytherm 8000 OC Instructions For Use Manual page 26

Resuscitation/open care unit
Table of Contents

Advertisement

Heat Therapy
Radiant Heater
Skin temperature control
The set temperature can be changed in 0.1 °C stages:
1 Keep p button or P button pressed until the
required temperature appears in the display
O
2 The yellow LED
is lit or flashes in step with the
control pulse (functional display of heating).
To test functioning during operation:
3 Press 6 button.
4 The simulated temperature value must be
36 °C (±0.1 °C) (function test).
If the sensor plug is disconnected or the sensor is
faulty:
5 Three flashing dashes in the middle of the display,
6 the red alarm LED Sensor is lit.
A continuous sound is given (cannot be switched off),
then:
re-connect sensor plug immediately or replace skin
temperature sensor.
If the fault cannot be immediately eliminated:
switch to »Man« operating mode.