Dräger Babytherm 8000 OC Instructions For Use Manual page 16

Resuscitation/open care unit
Table of Contents

Advertisement

Heat Therapy
Mattress Heating
Operational readiness
Switching unit on and activating self-test
1 Press the mains switch until it engages = ON.
The microprocessor self-test checks the unit for
proper functioning.
All LEDs lit. The numeric display shows 88.8 and an
alarm tone sounds.
2 After approximately 2 seconds, the unit shows the
standard specified temperature of 37.0 °C (display
blinks).
3 The display alternates periodically between SEt and
measured temperature.
4 Press button p or P briefly to adopt the specified
value.
3 Display continuously indicates the measured mattress
temperature.
5 Whenever the display alternates between Acc and
the measured temperature, and a tone sounds
periodically, the emergency storage battery is not
charged sufficiently.
The storage battery will be fully charged after
approximately 30 minutes of operation. The display
shows the measured temperature continuously, and
the tone does not sound.
To switch off the tone:
6 Press the button G, its yellow LED is lit.
The tone can be switched off for 15 minutes.
7 The display alternates between Err and the measured
temperature, and
8 the red LED Inop is lit – unit inoperative: see p. 37
for "Fault, Cause, Remedy".
16
D
3
°C
°C
>36 °C
>35 °C
2
D
3
°C
°C
>36 °C
>35 °C
4
4
D
5
°C
°C
>36 °C
>35 °C
6
D
8
7
°C
°C
>36 °C
>35 °C
Babytherm 8000
1
Check
Babytherm 8000
Check
Babytherm 8000
Check
Babytherm 8000
Check

Advertisement

Table of Contents
loading

Table of Contents