Endress+Hauser Proline Prosonic Flow B 200 Operating Instructions Manual page 13

Ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow B 200:
Table of Contents

Advertisement

Proline Prosonic Flow B 200
A
Vertical orientation
B
Horizontal orientation, transmitter head
up *
C
Horizontal orientation, transmitter head
down *
D
Horizontal orientation, transmitter head
at side
* A maximum deviation of only ±3 ° is permitted for the horizontal alignment of the
transducers.
Inlet and outlet runs
The sensor should be mounted upstream of assemblies such as valves, T-sections, elbows etc.
where possible. As a minimum, the inlet and outlet runs shown below must be observed to
achieve the specific accuracy of the device. The longest inlet run shown must be observed if two
or more flow disturbances are present.
For the dimensions and installation lengths of the device, see the "Technical Information"
document, "Mechanical construction" section
Endress+Hauser
Orientation
–3°
+3°
A0015545
A0015589
A0015590
A0015592
–3°
+3°
Mounting
A0015895
Compact version
A0016534
13