Do not touch the glass door and the inner surfaces for approx.10 hours after the prematurely
terminated sterilization cycle.
CO
control is deactivated during the sterilization process and will only be activated when
2
the temperature has cooled down to a value below 55
The chamber during the sterilization process can have negative impacts on the other chamber. Therefore it is recommended to
use the original stacking adaptor.
11.
Components of the CO
11.1.

Door lock

All INCU-Line units are equipped with a door lock located in the door handle.
Two keys are attached on the backside of the unit.
11.2.

Access ports

Two Ø30 mm access ports located on the left wall and on the rear can be used to insert external sensors, which have been
secured with silicone plugs. The plugs should cover the access ports while the unit is operating. If multiple cables have been
inserted through the access port and if it is not possible to use the plug, secure the access port with adhesive tape. If you leave
the access port open, it may affect temperature variation and fluctuation within the chamber.
51
vwr.com I Instruction manual VWR
incubator
2
®
CO
incubator
2
o
C.