Quick Change Of Parameters; Quick Change The Set Temperature - VWR INCU-Line Premium Instruction Manual

Co2 incubator
Table of Contents

Advertisement

10.3.

Quick change of parameters

You cannot make a quick change (time / temperature / CO
another user.
Information about the program owner can be found on the bottom panel
10.3.1.
In order to quickly change the value of the set temperature of the program, press the icon
The value of the temperature should be selected by scrolling the list up or down
.
Temperature cannot be higher than protection temperature (over temperature) -2°C.
29
vwr.com I Instruction manual VWR
Figure 19 – Main screen
Figure 20 - Starting the program

Quick change the set temperature

®
CO
incubator
2
) in a running program belonging to
2
(
Figure 6
in main screen
(Figure
21). Confirm the change by pressing
).
(Figure 2).

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents