2.
Performing Short Compensation
1.
For short compensation, we recommend using a short device with gold-plated
NOTE
surfaces, which provide stable contact resistance.
Figure 3-6
Performing Short Compensation
2.
Chapter 3
Press OPEN key to start the OPEN compensation data measurement. When the OPEN
compensation data measurement is completed, the softkey label OPEN on OFF (if it
is off) changes to OPEN ON off.
Put the probe adapter into the SHORTed state as shown in
appropriate device for shorting since no short device is supplied with the BNC
adapter.
Press SHORT key to start the SHORT compensation data measurement.When the
SHORT compensation data measurement is completed, the softkey label
SHORT on OFF (if it is off) changes to SHORT ON off.
Operation
Fixture Compensation
Figure 3-6
Use an
.
23