Appendix

Error code
System errors
40
Max. pulse frequency exceeded,
scaled pulse output, pulse factor
(5 kHz)
41
Min. pulse frequency below limit
< 0.00016 Hz
42
Entry > 2000 ms pulse width
43
Entry < 0.1 ms pulse width
44
Entry > 5.0 g/cm
45
Entry < 0.01 g/cm
46
Input too large
54
Zero flowmeter primary > 50 Hz
56
Entry > 3000 threshold
Detector empty pipe
74/76
Entry > 130% MAX – or MIN
alarm
91
Data in internal EEPROM invalid
92
Data in ext. EEPROM incorrect
93
Ext. EEPROM is incorrect or not
installed
94
Ver. ext. EEPROM incorrect
95
External flowmeter primary data
incorrect
96
Ver. EEPROM incorrect
97
Flowmeter primary incorrect
98
Ver. EEPROM is incorrect or not
installed
99
Input too large
Entry is too small.
7
Appendix
7.1

Additional documents

Operating Instructions (D184B132U02)
Data Sheet (D184S075U02
32 - EN
Error removal
Reduce pulse factor.
Increase pulse factor.
Decrease input value.
Increase input value.
3
density
Decrease input value.
3
density
Increase input value.
Reduce pulse width entry value.
Check the ground and ground signals. Adjustment can be made
if the flowmeter primary is filled with fluid and the flowrate is
zero.
Reduce entry value, check "Detector empty pipe" adjustment.
Decrease input value.
Data in internal EEPROM invalid, for corrective measures, see
Error code 5.
Data (e.g., R, damping) in external EEPROM is invalid, access
possible. Occurs when function "Store data in ext. EEPROM"
was not called. Use the function "Store data to ext. EEPROM"
to clear the error message.
No access possible, component defective. If the component is
not installed, the current ext. EEPROM that belongs to the
flowmeter must be plugged in above the display.
The database has not been updated to the present software
version. Use the function "Load data from ext. EEPROM" to
automatically update the external data. Use the function "Store
data to ext. EEPROM" to clear the error message.
See error code C.
Database version in the EEPROM is different from the installed
software. Clear the error by selecting "Update".
The data for the flowmeter primary in the internal EEPROM is
invalid. Clear the error by selecting "Load primary". (see error
code C)
No access possible, components defective. If the component is
not installed, the current EEPROM that belongs to the
flowmeter must be plugged in.
Reduce input value.
Increase input value.
)
FXE4000 (COPA-XE/MAG-XE)
D184B133U02