ABB FXE4000 Commissioning Instructions page 59

Electromagnetic flowmeter
Hide thumbs Also See for FXE4000:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 35
2. Current output (menu item "Current output")
Select the desired current range (0 ... 20 mA or 4 ... 20 mA)
3. For devices with a fieldbus, the bus address must be set (menu item "Data Link").
4. Pulse output (menu items "Pulse factor" and "Unit").
To set the number of pulses per volume flow unit, a unit for the totalizer (e.g., m
be selected under "Unit". Afterward the number of pulses has to be entered in the menu item
"Pulse factor".
5 Pulse width (menu item "Pulse width")
The pulse width at terminals V8 and V9 can be set between 0.1 ms and 2,000 ms.
6 System zero adjustment (menu item "System zero adjustment")
The fluid in the flowmeter primary must be at absolute zero. The flowmeter primary must be
full. Select the menu "System zero adjustment". Next press ENTER. Use the STEP key to
call up "automatic" and select ENTER to start the adjustment. During the automatic
adjustment, the flowmeter primary counts from 255 to 0 in the second display line. The
system zero point adjustment is completed when the counter reaches zero. The adjustment
lasts approx. 20 seconds.
D184B133U02
Nominal size
3
4
6
8
10
15
20
25
32
40
50
65
80
100
Nominal size
125
150
200
250
300
350
400
450
500
600
700
800
900
1000
FXE4000 (COPA-XE/MAG-XE)
Flow range end value
minimum (0.5 m/s)
0.2 l/min
0,4 l/min
1,0 l/min
1,5 l/min
2,25 l/min
5 l/min
7,5 l/min
10 l/min
20 l/min
30 l/min
3
3 m
/h
3
6 m
/h
3
9 m
/h
3
12 m
/h
Flow range end value
minimum (0.5 m/s)
3
21 m
/h
3
30 m
/h
3
54 m
/h
3
90 m
/h
3
120 m
/h
3
165 m
/h
3
225 m
/h
3
300 m
/h
3
330 m
/h
3
480 m
/h
3
660 m
/h
3
900 m
/h
3
1200 m
/h
3
1350 m
/h
Startup Operation
maximum (10 m/s)
4 l/min
8 l/min
20 l/min
30 l/min
45 l/min
100 l/min
150 l/min
200 l/min
400 l/min
600 l/min
3
60 m
/h
3
120 m
/h
3
180 m
/h
3
240 m
/h
maximum (10 m/s)
3
420 m
/h
3
600 m
/h
3
1080 m
/h
3
1800 m
/h
3
2400 m
/h
3
3300 m
/h
3
4500 m
/h
3
6000 m
/h
3
6600 m
/h
3
6900 m
/h
3
13200 m
/h
3
18000 m
/h
3
24000 m
/h
3
27000 m
/h
3
or l) must
EN - 25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fxe4000 copa-xeFxe4000 mag-xe

Table of Contents