When entering data, the transmitter remains online, i.e., current and pulse outputs show the
current operating mode. The functions of the individual keys are explained in the following:
ENTER function for magnet stick operation
The ENTER function is initiated when the DATA/ENTER sensor is activated for more than 3
seconds. The display flashes to indicate that the function is active.
There are two different methods of entering data:
Numeric entry
Entry from specified table
Note
When entering data, the values are checked for plausibility and, if necessary, rejected with an
appropriate message.
D184B133U02
C/CE
Toggle between operating mode and menu.
The STEP key is one of two arrow keys. Use STEP to scroll forward
STEP
through the menu. All the desired parameters can be called up.
The DATA key is one of two arrow keys. Use DATA to scroll backward
DATA
through the menu. All the desired parameters can be called up.
ENTER
The ENTER function requires that both arrow keys, STEP and DATA, be
pressed simultaneously. ENTER has the following functions:
Turn on/off Program protection.
Access the parameter to be changed and set the new, selected or
default parameter.
The ENTER function is only active for 10 seconds. If no entries are made
during this period, the old value is displayed on the transmitter.
FXE4000 (COPA-XE/MAG-XE)
Parameterization
EN - 27