2. Release the keys. The display will show:
The number shown (93.2 in this example) represents the gloss value
of the calibration tile stored in the gauge memory.
3. Press the READ/SELECT key. The display will show:
4. Use the SCROLL UP/DOWN key to adjust the value until it matches
the value quoted on the white calibration tile.
5. When the displayed value matches that shown on the tile, press the
READ/SELECT key to store the value. The gauge will show a confirmation message briefly and will
then return to the measurement mode.
Note: After the calibration value has been adjusted, the gauge must be re-calibrated before it is used for
gloss measurements - see "Calibration" on page 14.