R
5. Place the gauge carefully on the black calibration tile.
6. Press the READ/SELECT key.
After a short delay during which a 'Please Wait' message appears,
the display will show:
7. Place the gauge carefully on the white calibration tile.
8. Press the READ/SELECT key.
After a short delay during which a 'Please Wait' message appears,
the display will show:
The gauge will then return to the measurement mode and is ready
for use.
Note: In the Automatic Calibration Mode the Elcometer 6014 has an
automatic standard tile check routine. If the calibration tile is dirty or
damaged and the value does not conform to that stored in the memory the
display will show the following warning messages:
The white tile must be checked for permanent marks and deep scratches.
If these are seen on the tile the CANCEL key should be pressed - the
calibration value will not be stored.
16
Calibrate
Sample White Ref
45/0 Angle
Completed
Possible Error
Check Calib Tile
Accept New
Calibration?