6. Place the gauge carefully on the black calibration tile.
7. Press the READ/SELECT key.
After a short delay during which a 'Please Wait' message appears,
the display will show:
8. Place the gauge carefully on the white calibration tile.
9. Press the READ/SELECT key.
After a short delay during which a 'Please Wait' message appears,
the display will show:
The number shown (99.7 in this example) represents the value
measured for the tile. If the value does not match the certified value
of the calibration tile, use the SCROLL UP/DOWN key to adjust the
reading until it matches the value.
10. Press the READ/SELECT key to store the value and the gauge will
display:
The gauge will then return to the measurement mode and is ready
for use.
18
Calibrate
Sample High Ref
ANGLE CALIBRATE
45/0
99.7
45/0 Angle
Completed