The tile must then be carefully cleaned and inspected (see "Maintaining
your calibration tiles" on page 31). The calibration procedure must then
be repeated.
A new calibration tile will be required if permanent marks are present.
7.4 CALIBRATING IN MANUAL MODE
The manual calibration procedure is as follows:
1. Visually inspect the calibration tiles for smears or defects before you start.
2. Switch on the gauge and wait until the gauge is ready to take
measurements.
3. Use the SCROLL UP/DOWN button to select:
4. Press the READ/SELECT key. The display will show:
5. Press the READ/SELECT key. The display will show:
ARE YOU SURE ?
Press Read Key
CALIBRATE ANGLE
Press Read Key
ARE YOU SURE?
Press Read Key
Calibrate
Sample Zero Ref