Changing The Led - Labomed eVO 300 User Manual

Slit lamp
Table of Contents

Advertisement

eVO 300
4. Pull the bulb from the unit and replace the bulb
with the correct bulb as indicated in the specifications
section of the manual.
Note: Position the Bulb so the Cut Out in the metal
collar of the Bulb lines up with the Notch
in the lamp housing. Refer to Figure-24.
7. Move the Retaining Clamp back into its original
position. Refer to Figure-22.
8. Place the bulb holder back on the position.
9. Close the bulb door.
A

CHANGING THE LED

Figure 24
(For LED illumination only)
WARNING: NEVER REMOVE A LED THAT HAS RECENT-
LY BEEN IN USE AS IT MAY BE VERY HOT. WAIT UNTIL
IT IS COOLED.
WARNING: NEVER TOUCH A LED WITH BARE HANDS AS
FINGERPRINTS WILL SHORTEN THE LED LIFE.
1. Remove input power to the instrument.
2. Remove the Circlip by using circlip plier. shown as ''A''
B
Refer Figure 24.
Figure 25
3. Lift up the illumination housing carefully.
4. Remove the Lamp Cover by unscrewing the Four
Screws by using screwdriver shown as ''B''. Refer
Figure 25.
5. Unscrew the two M3 allen screws to disassemble the
Heat Sink by using 2.5 Allen wrench shown as ''C''. Refer
C
Image 26.
Figure 26
6. Unscrew the LED from Heat Sink by unscrewing the
Two Screws shown as ''D''. Refer to Figure 27.
D
7. Replace New LED and follow Reverse to complete
assembly.
Figure 27
16
eVO 300
Issue 1.1
8126200-795
Printed on April,2016