FUSE REPLACEMENT

Replace the fuses in the Power Input Module with the
fuses indicated in the Specifications section of this
manual.
1.
Remove input power to the instrument.
2.
Press down on the top tab in the middle of the
Power
Input
Module
and gently pull out the Fuse Holder by gripping the
two small tabs. Refer to figures-25 & 26
3.
Open the Door to the Fuse Holder by pulling
it down. Refer to figure-25.
NOTE:
The Fuses will pop up when the door is open,
making removal easier.
4.
Install new fuses into the Fuse Holder that is
indicated in the Specification section of this manual.
5.
Install the Fuse Holder by closing the door, and
pushing the Fuse Holder back until it snaps into place.

Replacement of Chin Rest Light Indicator

1. Unscrew the LED cap refer fig. 27.
2. Pull out the defective LED and replace with new LED.
refer fig.27a. Thread in the LED cap back.
NOTE:
If New LED does not light up after switching on.
remove it and re-fix after changing its polarity.
LED Cap
Figure 27 Changing
Target Light LED
17
8126200-795
to
release
the
Fuse
Pull LED Bulb
Figure 27a Changing
Target Light LED
eVO 300
Notch
Holder,
Figure-24 Notches
Figure-25 Pull Out
Open
Door
Figure-26 Open Fuse Door
eVO 300
Cut Out
Pull
Out
Tabs
Issue 1.1
Printed on April,2016