Changing The Led - Labomed eVO 350 User Manual

Slit lamp
Table of Contents

Advertisement

CHANGING THE LED

(For LED illumination only)
WARNING: NEVER REMOVE A LED THAT HAS RECENTLY
BEEN IN USE AS IT MAY BE VERY HOT. WAIT UNTIL IT IS
COOLED.
WARNING: NEVER TOUCH A LED WITH BARE HANDS AS
FINGERPRINTS WILL SHORTEN THE LED LIFE.
1. Remove input power to the instrument.
2. Remove the Circlip by using circlip plier. shown as ''A''
Refer Figure 24.
3. Lift up the illumination housing carefully.
4. Remove the Lamp Cover by unscrewing the Four
Screws by using screwdriver shown as ''B''. Refer
Figure 25.
5. Unscrew the two M3 allen screws to disassemble the
Heat Sink by using 2.5 Allen wrench shown as ''C''. Refer
Image 26.
6. Unscrew the LED from Heat Sink by unscrewing the Two
Screws shown as ''D''. Refer to Figure 27.
7. Replace New LED and follow Reverse to complete
assembly.
8126300-795
eVO 350
eVO 350
A
Figure 24
B
Figure 25
C
Figure 26
D
Figure 27
Issue 1.2
Printed on June,2020
16