GUIDANCE TABLES (continued)
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The eVO 300 is intended for use in the electromagnetic environment specified below. The
customer or user of the eVO 300 should ensure that it is used in such an environment.
Immunity
Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
voltage
Dips/Dropout
IEC 61000-4-11
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
22
8126200-795
Table 202 – Guidance and Manufacturer's Declaration
Electromagnetic Immunity
All Equipment and Systems
IEC 60601
Compliance
Test Level
±6kv Contact
±6kv Contact
±8kv Air
±8kv Air
±2kv Mains
±2kv Mains
±1kv I/Os
±1kv I/Os
±1kv Differential
±1kv Differential
±2kv Common
±2kv Common
>95% Dip
>95% Dip
for 0.5 Cycle
for 0.5 Cycle
60% Dip
60% Dip
for 5 Cycles
for 5 Cycles
30% Dip
30% Dip
for 25 Cycles
for 25 Cycles
>95% Dip
>95% Dip
for 5 Seconds
for 5 Seconds
3A/m
3A/m
eVO 300
Electromagnetic
Level
Environment
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the R/H should be
at least 30%.
Main power quality should be that of a
typical commercial or hospital environment.
Main power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the eVO 300 requires continued
operation during power mains interruptions,
it is recommended that the eVO 300 be powered
from an uninterruptible power supply or battery.
Power frequency magnetic field should be
that of a typical commercial or hospital
Environment.
eVO 300
- Guidance
Issue 1.1
Printed on April,2016