Configuration

Note
If you calibrate using the ProMinent buffer set, generally
you do not have to change the configuration.
The following basic settings can be changed in the configu-
ration:
Automatic calibration (Calimatic
on or off (manual), buffer set for automatic
calibration
Three-point calibration on or off
To activate the configuration hold down cal with the meter
+
switched off and then press on/off.
on/off
cal
The menu items of the configuration menu are worked
through in sequence.
To change the setting of the respective menu item, press
Y
B
Y or B.
Pressing cal switches to the next menu item and stores
cal
the displayed settings.
Pressing meas exits the configuration menu at any time.
meas
The value last displayed and possibly hanged will then not
be saved.
By pressing the Y or B key, you can select whether you
Automatic
or manual
wish to calibrate with the Calimatic
calibration
ognition system or with manual entry of the individual
buffer values.
When calibrating with Calimatic
tion (AutCal on), you only have to enter the buffer set used
once in the configuration menu. The buffer values are
stored at the correct temperature. During calibration the
meter then automatically recognizes the buffer used (fac-
tory setting: automatic calibration on, ProMinent buffer
solution).
Operation
8
)
automatic buffer rec-
automatic buffer recogni-