Download Print this page

ECOTEST BDBG-09S Operating Manual page 42

Detecting unit of gamma radiation

Advertisement

Given BLOCK = 0
D7 D6 D5 D4 D3 D2 D1 D0
Low byte
High byte
Low byte
High byte
...
Low byte
High byte
Low byte
High byte
Low byte
High byte
DER0 (Low byte)
DER1
DER2
DER3 (High byte)
Byte
0 D6 0
0
0 D2 D1 D0
2
2
2
2
2
2
3
2
1
0
-1
-2
D7 X
X
X
S 2
Low byte
High byte
Byte
Serial #_0 (Low byte)
Serial #_1
Serial #_2
Serial #_3 (High byte)
Year
Month
Release version
Debug version
Pulse number in the channel
0
Pulse number in the channel
1
Pulse number in the channel
1022
Pulse number in the channel
1023
Obtained spectrum accumulation period, s.
DER,
fixed point number,
Least significant bit = 0.01 µSv/h
Statistical error of measurement
D0,D1 - self-testing results of the detecting unit
D0=1 - failure of the high sensitivity detector
D1=1 - failure of the low sensitivity detector
Reliable measurement result character
D2=0 - result is true
D2=1 – result is false *
D6=1 – measurements from GM counter
D6=0 – measurements from high sensitivity detector
2
2
Temperature, binary number
-3
-4
S=0- above-zero temperature
S=1- below-zero temperature
2
2
6
5
4
D7=0- normal operation of heat sensor
D7=1-failure of heat sensor
Pulses per second (integrated)
Device model. 0хDD – BDBG-09S
Serial # of the detecting unit
Firmware version
...
41