3.2.4.3.17 Perform operations 3.2.4.3.16 of the OM for gamma radiation DER from
H 
*
with the value of
0
3.2.4.3.18 Perform operations 3.2.4.3.16 of the OM for gamma radiation DER from
H 
*
with the value of
0
3.2.4.3.19 Perform operations 3.2.4.3.16 of the OM for gamma radiation DER from
source with the value of
3.2.4.3.20 Perform operations 3.2.4.3.16 of the OM for gamma radiation DER from
H 
*
with the value of
0
3.2.4.3.21 The detecting unit is acknowledged to have passed the testing, if the limit of main
relative error of measurement of each level of gamma radiation DER does not exceed 15+1/
H 
*
is a numeric value of gamma radiation DER, equivalent to µSv/h.
(
10
)
, where
3.2.4.4 Presentation of testing results.
3.2.4.4.1 Positive results of periodic testing shall be certified by the issuance of a certificate of
the established form in accordance with the legislation in the field of metrology and metrological
activity or by an entry in the "Certificate of acceptance" section in the logbook for the detecting unit.
3.2.4.4.2 The detecting units that do not meet the requirements of the testing procedure are not
allowed for manufacture and use, and get the certificate of inadequacy in accordance with the
legislation in the field of metrology and metrological activity.
(
10
)
= (800100) µSv/h.
(
10
)
= (80.1) mSv/h.
H 
*
(
10
)
= (8010) mSv/h.
0
(
10
)
= (800100) mSv/h.
16
137
Cs source
137
Cs source
137
Cs
137
Cs source
H 
*
(