Detecting unit of gamma radiation

Advertisement

To receive gamma radiation spectrum from the detecting unit there are two frames «Expert»
and «Expert1».
Frame «Expert».
The information display system should transmit the following sequence BLOCK in "Expert"
frame:
1. Frame "Expert" BLOCK = 9 in BDBG (query for the new spectrum accumulation)
2. Frame "Expert" BLOCK = 9 from BDBG (response to accumulation start)
3. Frame "Expert" BLOCK = 0 in BDBG (query 0 ÷ 127 spectrum channels)
4. Frame "Expert" BLOCK = 0 from BDBG (response 0 ÷ 127 spectrum channels)
5. Frame "Expert" BLOCK = 1 in BDBG (query 128 ÷ 255 spectrum channels)
6. Frame "Expert" BLOCK = 1 from BDBG (response 128 ÷ 255 spectrum channels)
7. Frame "Expert" BLOCK = 2 in BDBG (query 256 ÷ 383 spectrum channels)
8. Frame "Expert" BLOCK = 2 from BDBG (response 256 ÷ 383 spectrum channels)
9. Frame "Expert" BLOCK = 3 in BDBG (query 384 ÷ 511 spectrum channels)
10. Frame "Expert" BLOCK = 3 from BDBG (response 384 ÷ 511 spectrum channels)
11. Frame "Expert" BLOCK = 4 in BDBG (query 512 ÷ 639 spectrum channels)
12. Frame "Expert" BLOCK = 4 from BDBG (response 512 ÷ 639 spectrum channels)
13. Frame "Expert" BLOCK = 5 in BDBG (query 640 ÷ 767 spectrum channels)
14. Frame "Expert" BLOCK = 5 from BDBG (response 640 ÷ 767 spectrum channels)
15. Frame "Expert" BLOCK = 6 in BDBG (query 768 ÷ 895 spectrum channels)
16. Frame "Expert" BLOCK = 6 from BDBG (response 768 ÷ 895 spectrum channels)
17. Frame "Expert" BLOCK = 7 in BDBG (query 896 ÷ 1023 spectrum channels)
18. Frame "Expert" BLOCK = 7 from BDBG (response 896 ÷ 1023 spectrum channels)
19. Frame "Expert" BLOCK = 8 in BDBG (query of the obtained spectrum parameters)
20. Frame "Expert" BLOCK = 8 from BDBG (response with the obtained spectrum parameters,
end of the accumulated spectrum reception in a time specified in BLOCK = 8)
To receive the next fragments of the accumulated spectrum one should repeat the operations
from 3 to 20 inclusive, but so that a new query of the accumulated spectrum fragment took place no
sooner than 2 s after the previous Frame "Expert» BLOCK = 0 in BDBG.
Frame «Expert1».
The data display system should transmit to the detecting unit the following sequence BLOCK
in "Expert1" frame:
1. Frame «Expert1» BLOCK = 9 in BDBG (query for a new spectrum accumulation start)
2. Frame «Expert1» BLOCK = 9 in BDBG (reply about accumulation start)
3. Frame «Expert1» BLOCK = 0 in BDBG (query 0 ÷ 1023 spectrum channels)
4. Frame «Expert1» BLOCK = 0 from BDBG (reply 0 ÷ 1023 spectrum channels)
To receive the following fragments of accumulated spectrum it is necessary to repeat operations
from 3 to 4, but so that the new query for fragment of the accumulated spectrum occurred not more
than 2 s after the previous frame "Expert1" BLOCK = 0 in BDBG.