Service Schedule - Breas Clearway 2 Service Manual

Hide thumbs Also See for Clearway 2:
Table of Contents

Advertisement

WARNING! Safety Precautions
• Follow the safety precautions when working on the Clearway 2
• Do not work on the Clearway 2 with the case open and the
power supply connected, unless instructed to do so in this
manual
• Always use extreme caution when working on the Clearway 2
when connected to any power source with the case open
• Do not use explosive gases and / or fluids near the Clearway 2
• Make sure that all precautions to prevent electrostatic discharge
(ESD) have been taken

Service Schedule

An inspection of the Clearway 2 (described in this chapter) is
advised to be carried out at least every 12 months.
Interval
Every 12 months
Every 5 Years
Every 15000 blower hours
2210.2657 Rev 3.0
Procedure
Perform an external inspection
Change the air inlet filter material
Perform a function test
Perform an electrical safety test
Perform the 12 month checks,
plus:
Change the clock battery
Change the alarm battery
Perform an internal inspection
Clean the valve
Change the blower
Maintenance Instructions
NIPPY Clearway 2 Service Manual
8
.

Hide quick links:

Advertisement

Table of Contents
loading