Replacement Parts; Preparing For Inspection - Breas Clearway 2 Service Manual

Hide thumbs Also See for Clearway 2:
Table of Contents

Advertisement

Replacement Parts

The following parts should be available when performing service on the
Clearway 2.
Part
Air inlet filter White
Air inlet filter Grey
Alarm battery
Clock battery
Blower

Preparing for Inspection

Clean the case exterior using 70% alcohol cleaning wipes.
Inspection record
(see Inspection record template, Appendix 13.4)
Record the serial number
Record the blower hours run
Record / save the patient settings as required
Determine the level of service required
Labels and markings
Labels and markings should be attached and fully legible:
Front panel markings.
Information label on rear panel
2210.2657 Rev 3.0
Part number
2210.3001
2210.3000
3/V110H
002129
2210.U51D2
NIPPY Clearway 2 Service Manual
When required
12 months
As Required
5 Years
5 Years
15000 hours
Maintenance Instructions
10
.

Hide quick links:

Advertisement

Table of Contents
loading