6.14

IEEE-488 Parameters

The calibrator may optionally be fitted with an IEEE-488 GPIB interface. In
this case the user may set the interface address and termination within the
IEEE-488 parameter menu. This menu does not appear on instruments not fit-
ted with the interface. The menu is indicated by,
Press "SET" to enter the menu.
6.14.1

IEEE-488 Address

The IEEE-488 interface must be configured to use the same address as the ex-
ternal communicating device. The address is indicated by,
AddrESS
Press "SET" to access the address setting.
Add=22
Adjust the value with "UP" or "DOWN" and then use "SET" to set the address
to the displayed value.
Add=15
6.14.2

Termination

The transmission termination character can be set to carriage return only, line-
feed only, or carriage return and linefeed. Regardless of the option selected the
instrument will interpret either a carriage return or linefeed as a command ter-
mination during reception. The termination parameter is indicated with,
Press "SET" to access the termination setting.
EOS=Cr
Use "UP" or "DOWN" to change the selection.
EOS=LF
Use "SET" to save the new selection.
IEEE
IEEE-488 parameters menu
IEEE-488 interface address
Current IEEE-488 interface address
New IEEE-488 interface address
IEEE-488 termination
EOS
Present IEEE-488 termination
New termination selection
6 Controller Operation
IEEE-488 Parameters
37