Table of Contents

Advertisement

Table 3. Digital Interface Command Summary
Command Description
Display Temperature
Read current set-point
Set current set-point to n
Read scan function
Set scan function:
Turn scan function on
Turn scan function off
Read scan rate
Set scan rate to n degrees per
minute
Read temperature
Secondary Menu
Read proportional band setting
Set proportional band to n
Read cutout setting
Set cutout setting:
Set cutout to n degrees
Reset cutout now
Read heater power
(duty cycle)
Ramp and Soak Menu
Read number of programmable
set-points
Set number of programmable
set-points to n
Read programmable set-point
number n
Set programmable set-point num-
ber n to n
Read program set-point soak
time
Set program set-point soak time
to n minutes
Read program control mode
Command
Command
Format
Example
s[etpoint]
s
s[etpoint]=n
s=450
sc[an]
sc
sc[an]=on/of[f]
sc[an]=on
sc=on
sc[an]=of[f]
sc-of
sr[ate]
sr
sr[ate]= n
sr=5
t[emperature]
t
pr[op-band]
pr
pr[op-band]=n
pr=8.83
c[utout]
c
c[utout]=n/r[eset]
c[utout]=n
c=500
c[utout]=r[eset]
c=r
po[wer]
po
pn
pn
pn= n
pn=4
ps n
ps3
ps n = n
ps3=50
pt
pt
pt= n
pt=5
pc
pc
7 Digital Communication Interface
Returned
Returned
Example
set: 9999.99 {C or F}
set: 150.00 C
scan: {ON or OFF}
scan: ON
srat: 999.99 {C or F}/min srat: 10.0 C/min
t: 9999.99 {C or F}
t: 55.69 C
pb: 999.9
pr: 15.9
c: 9999 {C or F}
c: 620 C, in
p%: 9999
po: 1
pn: 9
pn: 2
ps n : 9999.99 {C or F}
ps1: 50.00 C
ti: 999
ti: 5
prog: {OFF or ON}
prog: OFF
Interface Commands
Acceptable
Values
Instrument
Range
ON or OFF
.1 to 100°C
Depends on
Configuration
Temperature
Range
1 to 8
1 to 8, Instru-
ment Range
0 to 500