Controller Configuration; Operating Parameters; High Limit - Fluke 9150 User Manual

Portable furnace
Hide thumbs Also See for 9150:
Table of Contents

Advertisement

To store the new setting press "SET". Press "EXIT" to continue without storing
the new value.

8.9 Controller Configuration

The controller has a number of configuration and operating options and
calibration parameters which are programmable via the front panel. These are
accessed from the secondary menu after the proportional band function by
pressing "SET". Pressing "SET" again enters the first of three groups of
configuration parameters—operating parameters, serial interface parameters and
calibration parameters. The groups are selected using the "UP" and "DOWN"
keys and then pressing "SET".

8.10 Operating Parameters

The operating parameters menu is indicated by:

Press "SET" to enter the menu. The operating parameters menu contains the HL
(High Limit) parameter, the Soft Cutout parameter, and the Cutout Reset Mode
parameter.

8.10.1 High Limit

The HL parameter adjusts the upper set-point temperature. The factory default
and maximum are set to 1200. For safety, a user can adjust the HL down so the
maximum temperature set-point is restricted.

Press "SET" to enable adjustment of HL.

 
Adjust the HL parameter digit by digit using "UP", "DOWN", and "SET" as each
digit flashes.
Find Quality Products Online at:
Accept the new proportional band setting
Operating parameters menu
High Limit parameter
Access High Limit
Flashes the current value and then displays the value for
adjustment
Current HL setting
GlobalTestSupply
www.
Controller Operation
8.9
Controller Configuration
.com
sales@GlobalTestSupply.com
8
8-11