bration parameters. The groups are selected using the "UP" and "DOWN" keys
and then pressing "SET".
8.10

Operating Parameters

The operating parameters menu is indicated by:
Press "SET" to enter the menu. The operating parameters menu contains the
HL (High Limit) parameter, the Soft Cutout parameter, and the Cutout Reset
Mode parameter.
8.10.1

High Limit

The HL parameter adjusts the upper set-point temperature. The factory default
and maximum are set to 1200. For safety, a user can adjust the HL down so the
maximum temperature set-point is restricted.
Press "SET" to enable adjustment of HL.
S
1200
1200.0
Adjust the HL parameter digit by digit using "UP", "DOWN", and "SET" as
each digit flashes.
1005.9
Press "SET" to accept the new temperature limit.
8.10.2

Soft Cutout

The Soft Cutout parameter is used by the controller to shut the unit down dur-
ing over-temperature conditions.
SoFtCo
Press "SET" to enable adjustments of the Soft Cutout.
S
Par
Operating parameters menu
High Limit parameter
HL
Access High Limit
Flashes the current value and then displays the value for
adjustment
Current HL setting
New HL setting
Soft Cutout parameter
Access Soft Cutout
8 Controller Operation
Operating Parameters
31