Fluke 9116A User Manual page 20

Table of Contents

Advertisement

14
6.
Reinstall the control HTPRT (6) and the cutout thermocouple (7) as
shown in Figure 2 on page 13. They must be inserted far enough to clear
the top cover. They should also be centered in the air space between the
element and the heat pipe. Small pieces of the fiber ceramic material can
be used to fix their positions. The wires from the probes must be prop-
erly routed and connected to the controller as shown in Figure 2. Re-
move the cover over the electronics panel. Route the probe wires through
the hole in the top of the furnace (at the front right of the furnace when
facing the furnace) and down to the access hole into the electronics com-
partment (at the center left when facing the electronics panel). Connect
the probe wires to the controller. The labels on the probes and the con-
troller must be used to correctly complete the wiring. The thermocouple
wires should be firmly attached using the screw-down terminals on the
controller printed circuit board. The HPRT connection must be firmly
seated on the controller Make sure that the leads from the thermocouples
are directed away from the central well of the furnace, and will not be
pinched or interfere with other parts. See Figure 3
7.
Reinstall the top cover. Locate the cover over the ball catches and press
down firmly.

Advertisement

Table of Contents
loading

Table of Contents