Duplex Mode; Linefeed - Fluke 9116A User Manual

Table of Contents

Advertisement

9116A Furnace
User's Guide
6.13.3
6.13.4
36
Press "SET" to choose to set the sample period. The current sample period
value is displayed.
Current sample period (seconds)
SA= 1
Adjust the value with "UP" or "DOWN" and then use "SET" to set the sample
rate to the displayed value.
New sample period
SA= 60

Duplex Mode

The next parameter is the duplex mode. The duplex mode may be set to full du-
plex or half duplex. With full duplex any commands received by the calibrator
via the serial interface are immediately echoed or transmitted back to the device
of origin. With half duplex the commands are executed but not echoed. The du-
plex mode parameter is indicated by,
DUPL
Serial duplex mode parameter
Press "SET" to access the mode setting.
Current duplex mode setting
DUP=FULL
The mode may be changed using "UP" or "DOWN" and pressing "SET".
New duplex mode setting
DUP=HALF

Linefeed

The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return.
The linefeed parameter is indicated by,
LF
Serial linefeed parameter
Press "SET" to access the linefeed parameter.
LF=On
Current linefeed setting
The mode may be changed using "UP" or "DOWN" and pressing "SET".
New linefeed setting
LF=OFF