9116A Furnace
User's Guide
Digital Interface Command Summary continued
Command Description
Set program control mode:
Start program
Stop program
Continue program
Read program function
Set program function to n
Configuration Menu
Probe Menu
Read R0 calibration parameter
Set R0 calibration parameter to n r[0]=n
Read ALPHA calibration
parameter
Set ALPHA calibration parameter
to n
Read DELTA calibration
parameter
Set DELTA calibration parameter
to n
Operating Parameters Menu
Set temperature units:
Set temperature units to Celsius
Set temperature units to
Fahrenheit
Read cutout mode
Set cutout mode:
Set cutout to be reset manually
Set cutout to be reset
automatically
Read approach setting
Set approach setting to n
degrees
Read stability
Set soak stability to n degrees
Serial Interface Menu
Read serial sample setting
44
Command
Command
Format
Example
pc=g[o]/s[top]/c[ont]
pc=g[o]
pc=g
pc=s[top]
pc=s
pc=c[ont]
pc=c
pf
pf
pf= n
pf=2
r[0]
r
r=100.324
al[pha]
al
al[pha]=n
al=0.0038433
de[lta]
de
de[lta]= n
de=1.45
u[nits]=c/f
u[nits]=c
u=c
u[nits]=f
u=f
cm[ode]
cm
cm[ode]=r[eset]/a[uto]
cm[ode]=r[eset]
cm=r
cm[ode]=a[uto]
cm=a
ap[proach]
ap
ap[proach]= n
ap=15
ts
ts
ts= n
ts=.1
sa[mple]
sa
Returned
Returned
Example
pf: 9
pf: 3
r0: 999.999
r0: 100.578
al: 9.9999999
al: 0.0038573
de: 9.99999
de: 1.46126
cm: {xxxx}
cm: AUTO
ap:9
ap:5
ts:9.9
ts:0.5
sa: 9
sa: 1
Acceptable
Values
GO or STOP or
CONT
1 to 4
98.0 to 104.9
.00370 to
.00399
0.0 to 2.9
C or F
RESET or
AUTO
0 to 20°C
.01 to 4.99°C