9116A Furnace
User's Guide
6.5.2
6.5.3
28
S
Save new setting

Set-points

The next parameters are the program set-points.
First set-point
1 25.00
Use the "UP" or "DOWN" buttons to select any of the set-points.
Third set-point
3 500.00
Press "SET" to be able to change the set-point.
C 500.00
Set-point value
Use "UP" and "DOWN" to change the set-point value.
New set-point value
C 565.00
Press "SET" to save the new set-point value.
The other set-points can also be set in the same manner. Once the set-points are
programmed as desired press "EXIT" to continue.
E
Continue to next menu function

Program Soak Time

The next parameter in the program menu is the soak time. The soak time is the
time, in minutes, that the furnace maintains each of the program set-points after
settling before proceeding to the next set-point. The duration is counted from
the time the temperature settles to within a specified stability. The stability re-
quirement can be set in the parameter menu as explained in Section 6.12.4. The
default is 0.1°C.
Soak time in minutes
Pt=15
Use the "UP" or "DOWN" buttons to change the time.
Pt=5
New soak time
Press "SET" to continue.
S
Save new setting