Authorized Service Centers - Fluke 9116A User Manual

Table of Contents

Advertisement

under the unit and carefully lifting at the same time. Never move the fur-
nace if it is hot.
• Air circulated through the gap surrounding the furnace core keeps the
chassis cool. DO NOT SHUT OFF THE FURNACE WHILE AT
HIGH TEMPERATURES. The fan will turn off allowing the chassis to
become hot. Alternatively, if used, the cooling water should remain on un-
til the furnace is cool.
• Once the unit has been taken to high temperatures (over 800°C), it takes
days for the unit to cool completely.
1.4

Authorized Service Centers

Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
Chao Yang District
Beijing 100004, PRC
1 Before You Start
Authorized Service Centers
5

Advertisement

Table of Contents
loading

Table of Contents