Quick Start

CAUTION: READ SECTION 6 ENTITLED BATH USE before placing the
bath in service. Incorrect handling can damage the bath and void the war-
ranty.
This section gives a brief summary of the steps required to set up and operate
the 7100 bath. This should be used as a general overview and reference and not
as a substitute for the remainder of the manual. Please read
carefully before operating the bath.
4.1

Unpacking

Unpack the bath carefully and inspect it for any damage that may have oc-
curred during shipment. If there is shipping damage, notify the carrier
immediately.
Verify that all components are present:
• 7100 Bath
• Access Hole Cover
• Controller Probe
• Users Guide
• Drain Elbow
• Report of Test
• Fill Hole Cover
• 9930 Interface-it Software and Users Guide (optional)
If you are missing any item, please call an Authorized Service Center (see Sec-
tion 1.3).
4.2

Semi-Hermetic Compressor

The semi-hermetic compressor(s), located below the bath, have had their
mounting hardware tightened down to avoid damage during shipment. In addi-
tion, shipping foam has been placed between some of the refrigeration lines for
the same purpose. Upon installation, please remove all pieces of the shipping
foam and adjust the compressor's hardware for proper vibration absorption as
illustrated below in Figure 2.
4.3

Set Up

Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with fluid, installing the probe and connecting power. Consult
4 Quick Start
Unpacking
Section5 through7