Quick Start; Unpacking; Set Up; Power - Fluke 6050H User Manual

Calibration bath
Hide thumbs Also See for 6050H:
Table of Contents

Advertisement

4

Quick Start

CAUTION: READ SECTION 6 ENTITLED BATH USE before placing the
bath in service. Incorrect handling can damage the bath and void the
warranty.
This section gives a brief summary of the steps required to set up and operate
the 6050H bath. This should be used as a general overview and reference and
not as a substitute for the remainder of the manual. Please read Section 5
through 8 carefully before operating the bath.
4.1

Unpacking

Unpack the bath carefully and inspect it for any damage that may have oc-
curred during shipment. If there is shipping damage, notify the carrier
immediately.
Verify that all components are present:
• Bath
• Access Hole Cover
• Controller Probe
• Manual
• Report of Test
If you are missing any item, please contact an Authorized Service Center.
4.2

Set Up

Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with salt, installing the probe and connecting power. Consult Section 5
for detailed instructions for proper installation of the bath. Be sure to place the
bath in a safe, clean and level location. Remember that the bath operates at very
high temperatures.
Fill the bath tank with salt. (See Section 8.1.2, Heat Transfer Salt.) Refer to
sections 6 through 7 for more information on bath setup and use.
The control probe must be inserted through the lid into the bath and plugged
into the socket at the back of the bath. DO NOT operate the bath without the
control probe properly installed.
4.3

Power

Plug the bath power cord into a mains outlet of the proper voltage, frequency,
and current capability. See Section 3.1, Specifications on page 11, for power
4 Quick Start