G – Operating Gravity
This is the gravity value of the location where the scale will be operated.
With SETUP, USE displayed the current setting for the Operating Gravity parameter will be
shown. If the setting displayed is acceptable, press the UNITS key to proceed to the next
parameter.
Range is from 9.7000 to 9.8999. Note that the second digit can only be a 7 or 8.
Otherwise, using the numeric keys, enter a new value
next digit and then press the UNITS key to save it and proceed to the next parameter.
*
Consult Tech Support for the operating location gravity values.
CAL1 – Calibrate Local Scale
With SETUP, CAL1 displayed the current setting for the Scale Calibration Setting
parameter will be shown. If the setting displayed is acceptable, press the UNITS key to
proceed to the next parameter.
Otherwise, use the TARE key to toggle between the selections and then press the UNITS
key to save it and proceed to the next parameter. The available settings are:
=
Preform scale calibration
On
=
Skip scale calibration
OFF
To begin calibration, select On and then press the UNITS key. The display will change to
show SETUP, CAL10, 107378 the Scale Zero Count.
CAL10 – Local Scale Zero Count
With SETUP, CAL10 displayed the current setting for the Scale Zero Count will be shown.
SETUP
CAL1-0
107374
Press the UNITS key again. The display will change to show ------ (dashes) for a few
seconds and then change to the prompt for the calibration weight
SETUP
CAL1-0
------
20
NOTE: The Zero Count shown is an example. Each scale will
have a different zero count.
8526-M260-O1 Rev C C Series Operation
*
. Use the TARE key to move to the
SETUP
CAL1-U
060Lb